Finite element analysis of a TAVI implant: modelling of self-expandable devices

A calibration of the Nitinol constitutive parameters of self-expandable devices based on radial force tests is proposed. Optimized material properties were derived for each device, and accurate Finite Element analyses on  patient-specific models were conducted to compare results using standard literature versus calibrated parameters, highlighting the impact of the metallic frame properties on the simulation outcomes.

In particular, we propose optimal material parameters of the well-known and commonly adopted Nitinol constitutive model. We proved that using unprecise material parameters may induce inaccurate predictions inducing significant underestimation/overestimation of stress/strain patterns.

Post-implant device deformation and evaluation of stress patterns
Finite element simulation of a TAVI implant in a patient-specific model
Radial force and wall stress after a TAVI implant device

Goal: Calibration of optimal Nitinol constitutive parameters for a Finite element analysis of a TAVI implant in a patient-specific model and evaluation of stress/strain patterns

Methods: Modelling of patient-specific geometries, finite volume analysis and Nitinol constitutive parameters

Results: TAVI simulation with optimal material parameters permits a significant improvement in the prediction of the implant and estimation of stress/strain patterns